EasyManua.ls Logo

Atom Incu i 101 - Page 78

Atom Incu i 101
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
76
y “Humidity sensor condensing.”
NO
NO
NO
YES
YES
YES
Condensation has formed on the
humidity sensor.
The cable of the sensor module is
connected properly to the sensor
module connecting port.
Dry the humidity sensor.
Connect the connector of the cable
of the sensor module securely.
The alarm message is still
displayed even after the sensor
module is replaced with a functional
one.
Failure of the sensor module
Failure of the detection board
Condensation may form inside the sensor module when it gets cooled rapidly after the incubator has been
operated with a high humidity setting. The sensor module will return to its normal operation when the
condensation disappears.

Table of Contents

Related product manuals