Troubleshooting
MAINTENANCE
87
@7 “Communication to the main body abnormal. (E1025)”
NO
The cable connecting the control
panel with the main body is
defective.
Failure of the control board
Replace the cable.
@8 “Internal system abnormal. (E10)”
Failure of the heater control board
@9 “Internal system abnormal. (E259)”
Failure of the detection board
#0 “Internal system abnormal. Power cycle the incubator. (E2)”
NO
NO
YES
The incubator operates normally
after it is power cycled.
The detection board and the control
board are connected properly.
The cable is intact.
Failure of the detection board
Connect the connectors securely.
Replace the cable.