Atom Incu i 101 - DISASSEMBLY AND REPLACEMENT PROCEDURES; Stand Component Replacement

Atom Incu i 101
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
MAINTENANCE
93
u The weight screen cannot be displayed.
YES
NO
The weight screen display switch
appears on the screen.
See (2) e “The control panel does
not respond when the screen is
touched.”
The weight monitor function has not
been set yet or is set incorrectly.
Contact your local Atom
representative.
i The clock on the control panel does not operate properly.
NO
NO
YES
YES
The clock operates normally in normal
use, but when the unit is operated
after having been turned off for some
time, the clock shows the wrong time.
Failure of the LCD display board
Replace the button cell battery on
the LCD display board.
Replace the button cell battery on
the LCD display board.
The unit has been used for more
than four years or so.