54733-10BA2L
2 – 166 02.2017
1 System Description Volume 2
1.6 Detector Codes
The detector codes are used for automatically adjusting the operat-
ing point of the high voltage and have already been set by the man-
ufacturer. They only have to be checked or adapted when the elec-
tronics has been exchanged.
1 The count rates in the measuring range must be greater than 1000cps.
Detector type LB-No. Scintillator Detector code Restrictions
CrystalSENS LB 480-11
LB 480-12
50x50 NaI 0 with count rates in the mea-
suring range of >1000cps
LB 480-11
LB 480-12
50x50 NaI 1 with count rates in the mea-
suring range of <1000cps
LB 480-13
LB 480-14
40x35 NaI 0
1
LB 480-15
LB 480-16
25x25 NaI 0
1
LB 480-17
LB 480-18
44x5 NaI 2
1
UniSENS LB 480-2A
LB 480-2B
500 22
LB 480-2E
LB 480-2F
1000 13
LB 480-2I
LB 480-2J
1500 10
LB 480-2K
LB 480-2L
2000 10
SuperSENS LB 480-31
LB 480-32
150x150 23
TowerSENS LB 480-4. 1000 mm 31
2000 mm 32
3000 mm 33
4000 mm 34
5000 mm 35
6000 mm 36
7000 mm 37
8000 mm 38