EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB 480 - 2.23 Product Conditions

BERTHOLD TECHNOLOGIES LB 480
450 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 343
Volume 3 Menu Structure
3
4 ReadIn Time Define here the period of time over which the count rate is to be
read-in with ReadIn for each calibration point.
The statistical variation of the count rate is averaged over this time
period. The longer it is, the better the mean value. A time period of
30 s (default setting) is usually reasonable.
2.23 Product Conditions
Menu path: Device Config Setup Cal Parameter Product
Conditions.
On this menu you enter the basic product-specific settings that need
to be performed before calibration.
Entries for these parameters are required only in the following
cases:
For a 1-point calibration if full calibration is not possible.
For containers with a gas pressure as long as the calibration
cannot be performed at the same gas pressure as in operation.
For containers with varying gas pressure, if the gas pressure is
to be compensated for automatically.
For products containing natural radioactivity which is to be
compensated for automatically.
1 Measuring Path This menu item is required for Calibration Method 1-POINT to
determine the full count rate automatically.
The value to be entered is the length of the beam path through the
measured product.
100%
0%
100%
0%
0%
100%
Measuring Path
Point source
Rod detector
Product
Point detector
Measuring Path
Product
Rod source
Measuring Path
Point detector
Product

Table of Contents

Related product manuals