SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 361
Volume 3 Menu Structure
3
2.33 Digital Input
Menu path: Device Config Setup I/O Setup Digital
Input.
The digital input can be controlled by the process control system.
The reaction occurs when a short circuit of the input takes place;
the signal must be bounce-free.
IMPORTANT
If Safety Mode is enabled, the digital input is disabled automati-
cally.
The digital input is only available as an option. It is available for ver-
sions with HART signal, but not for detectors with fail-safe signal
output.
1 Digital In Function Here you can define the switching function of the digital input.
You have the following setting options:
OFF
The digital input is disabled.
HOLD
The measurement is frozen (hold mode) as long as the contact
is closed.
EMPTY
Runs an empty calibration. The input must be closed only
briefly in order to start the process.
FULL
Runs a full calibration. The input must be closed only briefly in
order to start the process.
2 Digital In State Indicates whether the input is open or closed.