SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 361
Volume 3 Menu Structure
3
2.33 Digital Input
Menu path: Device Config Setup I/O Setup Digital
Input.
The digital input can be controlled by the process control system.
The reaction occurs when a short circuit of the input takes place;
the signal must be bounce-free.
If Safety Mode is enabled, the digital input is disabled automati-
cally.
The digital input is only available as an option. It is available for ver-
sions with HART signal, but not for detectors with fail-safe signal
output.
1 Digital In Function Here you can define the switching function of the digital input.
You have the following setting options:
• OFF
The digital input is disabled.
• HOLD
The measurement is frozen (hold mode) as long as the contact
is closed.
• EMPTY
Runs an empty calibration. The input must be closed only
briefly in order to start the process.
• FULL
Runs a full calibration. The input must be closed only briefly in
order to start the process.
2 Digital In State Indicates whether the input is open or closed.