Menu Structure Volume 3
• 7-CHECK ERROR
The status of the calibration parameters could not be fully veri-
fied. In this case, repeat the calibration. If the error occurs
again, replace the detector, at least the detector electronics.
The above described plausibility check cannot prevent that the user
may read-in incorrect data. If in doubt, please consult a BERTHOLD
TECHNOLOGIES service technician.
The following settings are made automatically at Calibrate +
Preset:
– Compensation Mode = OFF
– Calibration Method = Multipoint
– Lower calibration point set to 0%
– Upper calibration point set to 100%
– Cal Curve Type = Normal
2.11 Sensor Configuration
Menu path: Device Config Setup Sensor Configuration.
The sensor settings are preset at the factory.
1 Date & Time Opens the menu for the date and time (page 3-323).
2 Sensor Settings Opens the menu for the detector code and HV settings (page 3-
324).
3 Config Slave Detectors Opens the menu for the configuration of the Slave detectors (page
3-327).