SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 341
Volume 3 Menu Structure
3
2.22 Cal Settings
Menu path: Device Config Setup Cal Parameter Cal
Settings.
The basic settings for the calibration are defined in this menu.
Changes in this parameter group have an effect only when you call
Calibrate (Device Config
Setup
Cal Parameter, page 3-
336).
1 Nuclide Enter here the isotope (nuclide) that is used in your source:
• Co-60
• Cs-137
• USER DEFINED
This entry controls the automatic decay compensation. The correct
entry is also important in a single-point calibration and for gas den-
sity compensation. The isotope used is listed on the type plate of
the shielding and on your delivery documents. The item USER
DEFINED allows you to use any isotope you want. In this case,
additional parameters are requested: Specify the half-life under
Half Life Time and the absorption coefficient at Absorption.
What happens when you have selected the wrong isotope?
As a result of the wrong decay compensation, you will get a deviat-
ing display only after several weeks or months, with the deviation
increasing over time. If you perform a calibration with the gas den-
sity compensation GDA, or a 1-point calibration, the wrong isotope
will also calculate wrong calibration data.