SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 359
Volume 3 Menu Structure
3
2.31 Level Threshold Settings
Menu path: Device Config Setup I/O Setup Digital
Output Level Threshold Settings.
1 Switch Function Here you define whether the switching function for the digital
output (open collector) is used as a High Alarm or Low Alarm.
The switching direction ensures the fail-safe function of the dig-
ital output (open collector).
You have the following setting options:
•MAX
Alarm is triggered if the limit value is exceeded (high alarm).
•MIN
Alarm is triggered if the limit value is not reached (low alarm).
2 Threshold Enter the level limit value in % where the alarm is to be output.
3 Hysteresis The hysteresis prevents switching back and forth caused by statis-
tical fluctuations of the measured values. The default value is 5%.