SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 363
Volume 3 Menu Structure
3
2.35 Service
Menu path: Device Config Setup Service.
This menu allows you to access various test functions, enter the
license key and perform a detector reset.
1 PMT Opens the menu showing the anode current of the photomultiplier
(PMT) (page 3-364).
2 Sensor Temperature Opens the menu showing the various detector temperatures (page
3-365).
3 Test Opens the menu with the various test functions (page 3-366).
4 Plateau Opens the menu for the plateau measurement and display of the
plateau values (page 3-371).
5 License Key The license key allows you to enable the detector for other applica-
tions (level, density). You can get a license key through your sales
partner or directly from BERTHOLD TECHNOLOGIES GmbH & Co.
KG.
Write down your license key in your operating manual. Each detec-
tor has its own license key.
6 Reset Device The menu offers you several ways to reset the detector or certain
functions:
• Active Detector No.
Choose the detector you want to reset.
– Detector No. = 0 selects the master detector.
– Detector No. = 1 selects slave 1.
–etc.
• SW RESET (Software Reset)
Starts the detector new. The function corresponds to the
switching off and on of the supply voltage.
• FACTORY RESET
Resets most settings to factory default. Some settings, such as
the adjustment of current output and the license key will be
preserved.