SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 363
Volume 3 Menu Structure
3
2.35 Service
Menu path: Device Config Setup Service.
This menu allows you to access various test functions, enter the
license key and perform a detector reset.
1 PMT Opens the menu showing the anode current of the photomultiplier
(PMT) (page 3-364).
2 Sensor Temperature Opens the menu showing the various detector temperatures (page
3-365).
3 Test Opens the menu with the various test functions (page 3-366).
4 Plateau Opens the menu for the plateau measurement and display of the
plateau values (page 3-371).
5 License Key The license key allows you to enable the detector for other applica-
tions (level, density). You can get a license key through your sales
partner or directly from BERTHOLD TECHNOLOGIES GmbH & Co.
KG.
Write down your license key in your operating manual. Each detec-
tor has its own license key.
6 Reset Device The menu offers you several ways to reset the detector or certain
functions:
Active Detector No.
Choose the detector you want to reset.
Detector No. = 0 selects the master detector.
Detector No. = 1 selects slave 1.
–etc.
SW RESET (Software Reset)
Starts the detector new. The function corresponds to the
switching off and on of the supply voltage.
FACTORY RESET
Resets most settings to factory default. Some settings, such as
the adjustment of current output and the license key will be
preserved.

Table of Contents