54733-10BA2L
3 – 438 02.2017
8 Error Handling Volume 3
8.4 Reset
The SENSseries detectors can be reset in various ways. Use the
menu Device Config Setup Service to perform a software
reset or reset all settings to factory default.
1. Software reset: SW RESET
Starts the new detector, the parameters remain unchanged.
The function corresponds to the switching off and on of the sup-
ply voltage.
2. Factory reset: FACTORY RESET
Resets most settings to factory default. Some settings, such as
the adjustment of current output and the license key will be
preserved. Turn the HART
®
Communicator off and on again
after you have initiated the reset.
Switching the supply voltage off/on also causes a reset similar to
the software reset, but in this case possible "hardware blocks" will
also be fixed. The parameters remain unchanged.
See also chapter 6.3, "Master Reset", page 3-408.
8.5 Operation Modes during Measurement
Select Operating Mode (menu Live Display, page 3-314) to
view the current operating status.
8.6 Error Reset
If an error or a warning is identified, an error message is displayed.
This error message must be reset. If the error still occurs after
resetting, it will be indicated again on the display.
If the error is corrected, but the error message is still displayed, the
error can be acknowledged as follows:
•with Acknowledge Error under Diagnostic Operating
Status
•with Software Reset under Device Config Setup
Service Reset Device
• by switching the supply voltage off and on.