Volume 3 5 Calibration
3
5 Calibration
Errors in the calibration or in the parameter setting can lead to
false results. This may possibly lead to loss of production, or to
damage in the system.
For testing, we recommend using the test settings in the Service
menu to simulate the calibration points.
A test run, ideally under operating conditions, will give you a high
level of safety that your calibration is accurate. The entire measur-
ing range has to be covered in this test run.
Basically we recommend to have commissioning carried out by the
BERTHOLD TECHNOLOGIES service.
Prerequisites for calibration with
the HART
®
Communicator
– The detector is installed correctly and is powered from the
mains (see Volume 2, chapter 2 and chapter 3).
–The HART
®
Communicator is connected to the current loop.