Menu Structure Volume 3
2.60 Error Status
Menu path: Diagnostic Operating Status Error Status.
1 Meas Setup Status Displays the status of the calibration parameters. The status is
updated after each decay compensation (at 09:01 h) and after each
date change. It informs you that the measurement is still working
safely. One of the following status messages is displayed:
– 0-OK
The calibration carried out is OK.
– 1-ERROR BACKGROUND
The count rate of the background radiation is higher than that
for the calibration points for empty or full (Cal Point 0% or Cal
Point 100%).
– 2-ERROR NOT SORTED
The calibration points are not entered in sequence and need to
be sorted.
– 3-ERROR CURVE NOT MONOTONOUS
Not all count rates are decreasing with increasing level.
– 4-ERROR MISSING CALIBRATION POINT
There is a calibration point with indication of the value in % but
without count rate.
– 5-COMPENSATION ERROR
A gas density compensation at exponential characteristic curve
is not possible.
– 6-DATE ERROR
A date is still set to the default value of 1.1.2000. At Calibrate,
check the Date parameter. At Restore, check the date in
Restore Date.
– 7-CHECK ERROR
The status of the calibration parameters could not be fully veri-
fied. In this case, repeat the calibration. If the error occurs
again, replace the detector, at least the detector electronics.