SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 327
Volume 3 Menu Structure
3
2.14 Config Slave Detectors
Menu path: Device Config Setup Sensor Configuration
Config Slave Detectors.
IMPORTANT
Date and time have to be checked before all other parameters and
updated, if necessary. If the device is more than a month out of
service, then an error message indicates that the date must be
updated.
1 Nbr of Slaves Number of slaves that are connected to and have communication
with the master.
2 Add Slave For multi-detector arrangements (cascading several detectors).
To add a slave detector, assign it a detector number between 1
to 16 and enter its device ID number.
3 Remove Slave Interrupt the communication to a slave by entering its detector
number.
4 Refresh Slave Table Updates the Detector Table.
5 Detector Table Shows all slaves in a clearly laid out table that are connected to the
master and communicate with the master. The table also shows the
function of the connected detector.
•MEAS
The detector is used for multi-detector operation (cascading).
•COMP
The detector is operated as compensation detector. For exam-
ple, for automatic gas density compensation.
The table also indicates whether an error message in a detector is
present.
If no error is present, then 0 is displayed in the relevant column.
An error message is displayed with a three digit number. The cause
of the error and suggestions for correcting the error are described
in "Error Handling" in chapter 8.