SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 375
Volume 3 Menu Structure
3
7 Compensation Mode Indicates with which additional compensation calibration was per-
formed.
•OFF
Any compensation is disabled (default).
•AUTO PRC
Compensation is enabled so that radiation from the product
does not distort the measured value.
•MANUAL GDA
The calibration has taken into consideration that the container
is under gas pressure during operation.
•AUTO GPC
Compensation is enabled so that fluctuations in the gas pres-
sure do not distort the measured value.
2.48 Meas Readings
Menu path: Device Config Meas Parameter Meas
Readings.
1 Level Shows the current level in %.
2 Cps Average Shows the current count rate averaged over the time constant.
3 Cps Live Displays the current count rate.
4 Cps Average Op Visible only if Compensation Mode = AUTO GPC or AUTO
PRC is selected.
Shows the compensated count rate which was used to calculate the
current measured value.
5 Cps Average Comp Visible only if Compensation Mode = AUTO GPC or AUTO
PRC is selected.
Displays the averaged count rate of the compensation detector.
6 Cps Sensor Comp Visible only if Compensation Mode = AUTO GPC or AUTO
PRC is selected.
Displays the non-averaged count rate of the compensation detec-
tor.
7 Active Detector No. Select the detector for which you want to display the non-aver-
aged count rate (Cps Single Detector).
Detector No. = 0 shows the count rate of the master detector.
Detector No. = 1 shows the count rate of slave 1.
•etc.
8 Cps Single Detector Shows the count rate of the detector, which you had selected with
Active Detector No..