Volume 3 Menu Structure
3
2.61 Log
This menu provides information about the history of the error mes-
sages and parameter changes. Up to 25 events can be entered.
1 Error Log Opens the menu displaying the errors that have occurred.
2 Modification Log Opens the menu showing the history of the settings made.
2.62 Error Log
Menu path: Diagnostic Log Error Log.
Update the error log by selecting Refresh Error Log before you
select one of the following menu items. Otherwise, no or only old
entries may be displayed.
Tip
In chapter 8 on page 3-429 you find a list of possible causes and
troubleshooting procedures.
1 Active Detector No. Choose the detector for which you want to show the current
entries or for which you want to display the pending error under
"Error Code".
Detector No. = 0 shows the error entries in the master detec-
tor.
Detector No. = 1 shows the error entries in slave 1.
•etc.
2 Refresh Error Log Updates entries in the error log table.
3 Error Log Table The last 25 errors are displayed in a table. The table contains the
error code and the time the error occurred. The prerequisite is that
the date and time were set correctly.
4 Reset Error Log Deletes all entries in the error log.
5 Error Code Indicates if an error message is present.
If no error is present, then "0" is displayed here.
An error message is displayed with a three digit number. The cause
of the error and suggestions for correcting the error are described
in chapter 8, "Error Handling".
5 Acknowledge Error Acknowledges the currently pending error.