Menu Structure Volume 3
– SHUTDOWN
An serious error is displayed and the following steps are carried
out:
– entry in the error log (the error is stored in the error mem-
ory)
– the error will be reported digitally via HART
®
– the error is reported binary via the digital output (Digital
Out) if this output has been set to WARNING + ERROR
(see chapter 2.30, page 3-357)
– the fault current goes to >21mA
– the measurement is stopped.
The error remains active until it is eliminated and a reboot or
software reset has been carried out. If the error does not disap-
pear after restart (power off and then on, or software reset,
page 3-363), then the detector must be replaced.
– HOLD
The measurement is in the hold state, which means that the
measured value and the current output signal are frozen. This
status can occur under the following conditions:
– the plateau recording is running
– interference radiation is detected
– the digital input (option) is shorted
– TEST
A test value has been enabled in the menu Service Test (page
3-366).
3 Error Code Indicates if an error message is present.
If no error is present, then 0 is displayed here.
An error message is displayed with a three digit number. The cause
of the error and suggestions for correcting the error are described
in chapter 8, Error Handling.
5 Acknowledge Error Acknowledges the currently pending error.