54733-10BA2L
3 – 386 02.2017
Menu Structure Volume 3
SHUTDOWN
An serious error is displayed and the following steps are carried
out:
entry in the error log (the error is stored in the error mem-
ory)
the error will be reported digitally via HART
®
the error is reported binary via the digital output (Digital
Out) if this output has been set to WARNING + ERROR
(see chapter 2.30, page 3-357)
the fault current goes to >21mA
the measurement is stopped.
The error remains active until it is eliminated and a reboot or
software reset has been carried out. If the error does not disap-
pear after restart (power off and then on, or software reset,
page 3-363), then the detector must be replaced.
HOLD
The measurement is in the hold state, which means that the
measured value and the current output signal are frozen. This
status can occur under the following conditions:
the plateau recording is running
interference radiation is detected
the digital input (option) is shorted
TEST
A test value has been enabled in the menu Service Test (page
3-366).
3 Error Code Indicates if an error message is present.
If no error is present, then 0 is displayed here.
An error message is displayed with a three digit number. The cause
of the error and suggestions for correcting the error are described
in chapter 8, Error Handling.
5 Acknowledge Error Acknowledges the currently pending error.