SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 381
Volume 3 Menu Structure
3
2.54 Diagnostic
This menu provides status and error information and allows you to
view the error logs and setting changes logs (parameter changes).
Tip
In chapter 8 on page 3-429 you find a list of possible causes and
troubleshooting procedures.
1 Operating Status Opens the menu showing the operating status (page 3-382).
2 Log Opens the menu showing the logs for error and setting changes
(page 3-387).
3 Safety The status of all settings listed here must be OFF when you have
enabled Safety ON (Device Config Access, page 3-376).
4 Slave Status Shows on which slave there is currently an error.
A list of 16 slave detectors with error status is displayed.
OFF = error free
ON = an error is indicated
Tip
If an error is present, you can select the respective slave under
Diagnostic Error Log and you can read out the error status of
the respective detector.
5 Error Status 100 … 907 Here you can check the status of all error messages that are possi-
ble in the detector.
•OFF = error free
•ON = an error is indicated

Table of Contents