SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 337
Volume 3 Menu Structure
3
7 Recall Allows you to copy the current data set from Meas Parameter to
Cal Parameter. This allows you to revise the valid parameters,
without this having an influence on the measurement at first. After
you have finished editing, you can activate the changed settings
with Calibrate.
Recall is also necessary because the count rates in the calibration
parameters (Cal Parameter) in contrast to the count rates of the
measuring parameters (Meas Parameter) are not decay compen-
sated.
Recall overwrites all settings of the Cal Parameter menu.
2.21 Cal Points
Menu path: Device Config Setup Cal Parameter Cal
Points.
Changes in this parameter group have an effect only when you call
Calibrate (Device Config
Setup
Cal Parameter
Calibrate, page 3-335).
This menu allows you to adjust the measuring system.