SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 373
Volume 3 Menu Structure
3
2.45 Meas Parameter
Menu path: Device Config Meas Parameter.
These menus show the currently valid measurement parameters.
1 Meas Data Shows the menu with the currently measured values (page 3-
373).
2 Meas Settings Opens the menu showing the most important detector settings
(page 3-374).
3 Meas Readings Opens the menu presenting an overview of all raw count rates that
are specifically needed for a compensation measurement (page 3-
375).
2.46 Meas Data
Menu path: Device Config Meas Parameter Meas Data.
The menu shows the measured and automatically calculated mea-
surement settings (parameters). The count rates are corrected daily
at 09:01 h through the automatic decay compensation; the values,
therefore, become smaller in the course of a few weeks.
1 Background Displays the count rate for the background radiation (background)
which is present in the environment.
2 Meas Table After calibration, the calibration points are displayed here in a table.
Using this table, the measurement shows the current level.
3 Measurement Curve After calibration, the calibration points are displayed here in a cali-
bration curve.