SENSseries LB 480 Level measurement
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 315
Volume 3 Menu Structure
3
2 Level Indication of the current level in percent (primary variable).
The measured value shows you if
calibration utilizes or exceeds the measuring range
the calibration is correct
a possible gas density was sufficiently taken into account when
calibrating
3 Current Out Displays the actual output current at the analog output in mA.
4 Cps Average Op Visible only if Compensation Mode = AUTO GPC.
Displays the compensated count rate from which the measured
value is calculated (level in %) when AUTO GPC is enabled.
5 Cps Average Shows the current count rate averaged over the time constant.
6 Cps Live Shows the current non-averaged count rate.
7 Cps Average Comp Visible only if Compensation Mode = AUTO GPC
Displays the averaged count rate of the compensation detector.
8 GPC Comp Factor Visible only if Compensation Mode = AUTO GPC
Displays the compensation factor for the gas density compensation.
9 HV Live Displays the current HV (high voltage) at the photomultiplier. If HV
Mode is set to AUTO (normal operation), the values in HV Live
must change, viewed over several seconds.
10 Device Temp. Displays the temperature inside the detector.
11 Trend Displays the trend of important measured values such as level,
count rate and temperature.