54733-10BA2L
3 – 320 02.2017
Menu Structure Volume 3
2 Background The term background refers to the background radiation which is
present in the environment. The background radiation has to be
measured so that the detector can correctly compensate for the
decay of the radiation source.
Select ReadIn to read in the count rate for the background
radiation.
If you already know the value, you can enter it here.
3 Cal Point 0% Count rate with empty container or at a level below the measuring
range.
Select ReadIn to read in the count rate.
If you already know the value, you can enter it here.
4 Cal Point 100% Count rate with full container or at a level above the measuring
range.
Select ReadIn to read in the count rate.
If you already know the value, you can enter it here.