ICEfx™ Cryoablation System
WARNING. Use of cables other than those specied, with the exception of those sold by
Galil Medical for use as replacement parts for internal components, may result in increased
emissions or decreased immunity of the ICEfx Cryoablation System.
WARNING. The ICEfx Cryoablation System should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the ICEfx System should be
observed to verify normal operation.
WARNING. Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of
the ICEfx Compact Cryoablation System, including cables specied for use with the system.
Otherwise, degradation of the performance of this equipment could result.
Table 9-2. Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The ICEfx Cryoablation System is intended for use in a professional healthcare facility
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
The ICEfx System uses RF energy only for its internal
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11 Class A
Class A
Complies
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and
equipment.