8-2 User Manual
8.2 Electronics, Electrical and User Error Related Issues
Symptom Potential Causes / Solutions
System does not turn ON (i.e., the
fan is not running) or power is lost
during procedure
1. The power switch on the rear panel is OFF (Figure 2-2).
Turn the power ON.
2. The power cable to the ICEfx Console is disconnected
from the power outlet or from the rear panel of the
Console. Connect the power cable to the Console,
checking that the power cable is fully seated. Connect
the power cable to the power outlet.
3. There is no power at the wall outlet. Ensure the power
outlet is switched ON. Call the hospital Biomedical
Engineer for assistance, if needed.
4. A fuse may have blown. Reference Section 8.2.1 for
information on how to replace fuses in the system.
Channel or needle is not
recognized by the system and not
available for use
1. Check the corresponding Channel Locking Bar and
ensure it is fully in the locked position.
2. At least one needle must be inserted into a channel for
the channel to be available for use.
3. If steps 1 and 2 are unsuccessful, select a new needle
for use. Contact Galil Medical Customer Service to
arrange return of the non-operational needle.
4. The channel may be defective. Do not use this channel.
Move the needle(s) to a dierent channel. Conduct the
Needle Integrity and Functionality Test again.
USB ash drive is not functioning
Or
The USB ash drive is not
recognized by the system
1. The USB ash drive is not properly connected to the
USB port. Remove the USB ash drive from the USB
port on the Console. Wait a few seconds and reconnect
the USB ash drive to the USB port.
2. If the problem persists, try using a dierent Galil Medical
supplied USB ash drive.
3. The USB ash drive is faulty. Replace the USB ash
drive with a new Galil Medical supplied USB ash drive.
The Login Screen appeared
after the system was left idle for
more than 2 hours while on the
Procedure Screen
Enter the appropriate password to return to the Procedure
Screen.