EasyManua.ls Logo

BTG ICEfx - Reports Messages

BTG ICEfx
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ICEfx™ Cryoablation System
8-21
Message Reason for Appearance / Solutions
60-12 Unable to export logs to the USB ash drive
The USB ash drive may be disconnected or full.
Reason for Appearance:
During log export, an error occurred that
prevented completion of the export.
Solutions:
1. Verify that the USB ash drive is
connected and has enough space.
2. Contact Galil Medical Customer
Service if further assistance is
required.
8.8.6 Reports Messages
Message Reason for Appearance / Solutions
70-03 Report Error
Errors occurred while compiling the report
The report may be incomplete
Reason for Appearance:
User chose to access a report during a
procedure or to save data to a report at the
end of a procedure. Errors occurred which
may a󰀨ect the completeness of the report.
Solutions:
Contact Galil Medical Customer Service.
70-04 Report Saved
Report saved successfully
Reason for Appearance:
Report successfully saved to a USB ash
drive.
Solutions:
No user action is required.
70-05 Duplicate Filename
The chosen lename already exists on the USB
ash drive
Choose a di󰀨erent lename
Reason for Appearance:
User tried to export a report to a USB
ash drive using a lename duplicate to a
lename on the ash drive.
Solutions:
To export the report, choose a di󰀨erent
lename.

Table of Contents