EasyManua.ls Logo

BTG ICEfx - Multi-Point Thermal Sensors; Displayed Messages

BTG ICEfx
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-10 User Manual
8.7 Multi-Point Thermal Sensors
Symptom Potential Causes / Solutions
Not all of the expected
temperature sensor locations on
the MTS needle are registered on
the Procedure Screen.
1. On the Settings Screen, select the desired number of
sensors to display.
2. Move the MTS needle to a di󰀨erent port.
3. Replace the MTS needle.
4. Contact Galil Medical Customer Service.
MTS needle is connected but
temperature is not registered on
the Procedure Screen.
1. Move the MTS needle to a di󰀨erent port.
2. Replace the MTS needle.
3. Contact Galil Medical Customer Service.
8.8 Displayed Messages
The ICEfx System displays a message on the user interface when a user requests assistance or
when user, needle or system errors are detected.
NOTE: Record and report the message number (e.g., 10-01, 80-02) if assistance is required from
Galil Medical Customer Service.
8.8.1 Login Messages
Message Reason for Appearance / Solutions
10-01 Login
You have not entered the correct Login Name
Reenter your Login Name
Contact your System Administrator if assistance is
required
Contact Galil Medical Customer Service if further
assistance is required
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
No name was entered.
OR
The name entered did not match
assigned names in the system.
Solutions:
1. Reenter your login credentials.
2. Contact the System Administrator for
further assistance.
3. Contact Galil Medical Customer
Service.

Table of Contents