EasyManua.ls Logo

BTG ICEfx - Needle Messages (CX Needles)

BTG ICEfx
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-16 User Manual
8.8.4 Needle Messages (CX Needles)
Message Reason for Appearance / Solutions
40-01 Unsupported Needle
The connected needle is not supported by the
software
Choose and connect a di󰀨erent needle type
Reason for Appearance:
A needle type not supported by the
software conguration was connected in
the channel. The channel will be disabled
until an appropriate needle is connected.
OR
Software identied needles not regulatory
cleared or supported in the country of
operation. The channel will be disabled
until an appropriate needle is connected.
Solutions:
1. Disconnect and segregate the needle.
2. Connect a supported needle type
and select the needle type from the
drop-down menu.
3. Contact Galil Medical Customer
Service if further guidance is needed.
40-03 Recalled Needle
This needle lot number is identied as part of a
recall and is unavailable for use
Return the needle to Galil Medical
Connect a needle from a di󰀨erent lot number to
continue the procedure
Reason for Appearance:
The needle lot number was identied as
a lot number in a regulatory recall. The
channel will be disabled until a needle from
a di󰀨erent lot number is connected.
Solutions:
1. Disconnect and segregate the needle.
2. Return the needle to Galil Medical.
3. Connect a needle from a di󰀨erent lot
number to continue the procedure.
40-04 Expired Needle
The needle ‘Use by’ date is in the past
Replace with a new needle
Reason for Appearance:
The needle expiration date was identied
as an expired date. The channel will be
disabled until an appropriate needle is
connected.
Solutions:
1. Disconnect and segregate the needle.
2. Replace with a new needle.

Table of Contents