8-16 User Manual
8.8.4 Needle Messages (CX Needles)
Message Reason for Appearance / Solutions
40-01 Unsupported Needle
• The connected needle is not supported by the
software
• Choose and connect a di󰀨erent needle type
Reason for Appearance:
A needle type not supported by the
software conguration was connected in
the channel. The channel will be disabled
until an appropriate needle is connected.
OR
Software identied needles not regulatory
cleared or supported in the country of
operation. The channel will be disabled
until an appropriate needle is connected.
Solutions:
1. Disconnect and segregate the needle.
2. Connect a supported needle type
and select the needle type from the
drop-down menu.
3. Contact Galil Medical Customer
Service if further guidance is needed.
40-03 Recalled Needle
• This needle lot number is identied as part of a
recall and is unavailable for use
• Return the needle to Galil Medical
• Connect a needle from a di󰀨erent lot number to
continue the procedure
Reason for Appearance:
The needle lot number was identied as
a lot number in a regulatory recall. The
channel will be disabled until a needle from
a di󰀨erent lot number is connected.
Solutions:
1. Disconnect and segregate the needle.
2. Return the needle to Galil Medical.
3. Connect a needle from a di󰀨erent lot
number to continue the procedure.
40-04 Expired Needle
• The needle ‘Use by’ date is in the past
• Replace with a new needle
Reason for Appearance:
The needle expiration date was identied
as an expired date. The channel will be
disabled until an appropriate needle is
connected.
Solutions:
1. Disconnect and segregate the needle.
2. Replace with a new needle.