ICEfx™ Cryoablation System
7-1
7 SYSTEM CARE and MAINTENANCE
7.1 Cleaning the ICEfx™ Cryoablation System
Clean the ICEfx Cryoablation System after each use by following the steps below.
1. Clean the Touch Screen Monitor when the ICEfx Cryoablation Console is powered OFF.
• Wipe the screen gently with a damp gauze wipe.
• Use water or isopropyl alcohol cleaning solutions.
• Do not use cleaning agents such as Betadine or bleach solution.
2. Clean the Console by wiping with a damp gauze wipe.
• Use soap and water or isopropyl alcohol cleaning solutions.
• Do not use cleaning agents such as Betadine or bleach solution.
• Do not allow water or other liquid to drip or seep inside the needle connection ports. The
needle connection ports must remain completely dry at all times. Avoid system exposure to
liquids.
3. Ensure that cleaned surfaces are dry prior to closing or turning on the system.
CAUTION. Do not set food, drinks, or other objects on top of the Console. Doing so may damage
the system.
7.2 Service and Preventive Maintenance
Only trained and authorized service engineers are permitted to conduct service and system
preventive maintenance on the system. Preventive maintenance on the ICEfx Cryoablation System is
performed every two years.
CAUTION. Do not attempt to service the ICEfx Cryoablation System. If service is required, contact
Galil Medical Customer Service.
The ICEfx Cryoablation System will display a reminder on the screen approximately one month
prior to when the preventive maintenance is due. If the reminder message appears and the
preventive maintenance has not yet been scheduled, contact your Galil Medical authorized service
representative to schedule the service.
7.3 System Disposal
In compliance with Waste Framework Directives, when a ICEfx Cryoablation System is
decommissioned, contact Galil Medical Customer Service to schedule system replacement,
refurbishing or disposal.