ICEfx™ Cryoablation System
8-15
Message Reason for Appearance / Solutions
20-19 Procedure Timeout
The procedure has exceeded the allowable time
Procedure will be terminated
Reason for Appearance:
The procedure exceeded the allowable 8
hour duration.
Solutions:
If the message appeared because the
system was left unattended, shutdown
the system.
OR
If more time is required for the
procedure, enter the Procedure Screen
again and continue with the procedure.
8.8.3 Gas Messages
Message Reason for Appearance / Solutions
30-02 Vent System
The system was not vented at the end of the last
case
Vent the system
Reason for Appearance:
The user did not vent the system
completely before the last shutdown.
Solutions:
To clear the remaining gas pressure,
manually vent the gas using the Vent
Valve.
30-03 Low Gas Level
Low level of argon gas remains in the cylinder
Replace the gas cylinder as soon as feasible with
a new cylinder
Reason for Appearance:
Remaining gas in the cylinder was
calculated to last only X more minutes.
Where X represents the Low Cylinder
Alert set on the Settings Screen (reference
Section 4.8).
Solutions:
Replace the gas cylinder as soon as
possible.