Message Reason for Appearance / Solutions
20-19 Procedure Timeout
• The procedure has exceeded the allowable time
• Procedure will be terminated
Reason for Appearance:
The procedure exceeded the allowable 8
hour duration.
Solutions:
• If the message appeared because the
system was left unattended, shutdown
the system.
OR
• If more time is required for the
procedure, enter the Procedure Screen
again and continue with the procedure.
8.8.3 Gas Messages
Message Reason for Appearance / Solutions
30-02 Vent System
• The system was not vented at the end of the last
case
• Vent the system
Reason for Appearance:
The user did not vent the system
completely before the last shutdown.
Solutions:
To clear the remaining gas pressure,
manually vent the gas using the Vent
Valve.
30-03 Low Gas Level
• Low level of argon gas remains in the cylinder
• Replace the gas cylinder as soon as feasible with
a new cylinder
Reason for Appearance:
Remaining gas in the cylinder was
calculated to last only X more minutes.
Where X represents the Low Cylinder
Alert set on the Settings Screen (reference
Section 4.8).
Solutions:
Replace the gas cylinder as soon as
possible.