ICEfx™ Cryoablation System
8-5
Instruction Photo
9. Plug in the power cord.
10. Call Galil Medical Customer Service to schedule
a service call to determine the cause of the blown
fuses, to determine if service is required, and to
order replacement fuses.
8.3 Gas Issues
Symptom Potential Causes / Solutions
The ICEfx System does not allow testing
of a needle in a locked channel
1. Verify that the system is at operating pressure.
Reference Table 3-2 for information on working gas
pressures.
2. The Argon Shuto󰀨 Valve may be in the OFF
position. Verify that the Argon Shuto󰀨 Valve (Figure
2-2) is in the ON position to allow su󰀩cient gas
ow.
3. If necessary, further open the argon cylinder valve
by turning the valve counterclockwise to allow
su󰀩cient gas ow. Verify that appropriate pressure
is displayed on the Gas Indicator.