6-2 User Manual
5. Press the Shutdown button on the Login Screen to power down the system. A message appears
requesting conrmation to shut the system down.
6. Wait until the screen goes black. Turn the Power Switch on the rear of the Console to the OFF
position.
7. Unplug the Console.
WARNING. Do not pull on the power cord. Grasp the plug, not the power cord, to
disconnect the device from the wall socket.
8. Clean the Console after each use using the instructions provided in Section 7.1. Ensure the
system is dry prior to storage.
9. Lower the Touch Screen Monitor into the monitor storage compartment prior to system storage.
CAUTION. Prior to lowering the monitor, ensure that no objects are located in the monitor storage
compartment. Use care when lowering the monitor into the storage compartment; do not use
excessive force to avoid damaging the monitor.
CAUTION. Use care when lowering the touch screen monitor to avoid the potential for pinching
ngers.
6.2 Removing the Console from the Cart
The following procedure describes how to detach the Console from the cart if required for storage, or
if the Console is being returned to Galil Medical for service.
1. Loosen the retaining screw that secures the cart to the Console.
2. Lift the Console o the cart and transfer to a stable surface.