ICEfx™ Cryoablation System
8-25
Message Reason for Appearance / Solutions
80-35 System Error
The gas ow rate in channel XX exceeds the
system operation requirements and may impact
performance. Reduce the number of active
needles.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
The calculated ow rate for a given
channel exceeded the recommended ow
rate for a channel. The number of active
needles should be reduced.
Solutions:
1. Reduce the number of active needles.
2. Contact Galil Medical Customer
Service for further guidance.
80-36 System Error
The gas ow rate exceeds the system operation
requirements and may impact performance.
Reduce the number of active needles
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
The cumulative ow rate for all channels
exceeded the recommended ow rate for
the system. The number of active needles
should be reduced.
Solutions:
1. Reduce the number of active needles.
2. Contact Galil Medical Customer
Service for further guidance.
8.8.8 Service Messages
Message Reason for Appearance / Solutions
90-01 Service Due
Service the ICEfx System soon
Service must be completed by [DATE].
Contact Galil Medical Customer Service to
schedule service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
User was reminded of the deadline to
complete scheduled system service.
Reminder begins four weeks prior to the
service due date.
Solutions:
Contact Galil Medical Customer Service to
schedule service.
90-02 Service Due
Service for the ICEfx System is past due
Contact Galil Medical Customer Service to
schedule service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
The ICEfx System was not serviced on
schedule. During subsequent start-ups, the
user is reminded the service is past due.
Solutions:
Contact Galil Medical Customer Service to
schedule service.