8-24 User Manual
Message Reason for Appearance / Solutions
80-31 System Error
Gas cylinder shut o󰀨 valve is not open enough
to provide adequate ow. If necessary, open the
valve approximately another half turn.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
Gas ow from the argon cylinder was
insu󰀩cient. To improve gas ow, the
cylinder valve should be opened further.
Solutions:
Check that the argon cylinder valve is
su󰀩ciently opened to allow gas ow
(reference Section 4.7). If necessary, open
the valve approximately another half turn.
80-32 System Error
Channel X is defective. Choose another channel.
At the conclusion of the procedure, contact Galil
Medical Customer Service.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
A defective solenoid on Channel X was
detected; another channel should be
selected.
Solutions:
1. Select a di󰀨erent channel.
2. After completing the procedure,
contact Galil Medical Customer
Service.
80-33 System Error
Defective MTS in channel [X]. Replace with a
new MTS.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
The MTS temperature display was out
of the expected range when initially
connected to the needle connection panel.
Solutions:
1. Disconnect and segregate the MTS
2. Connect a new MTS
80-34 System Error
Fan X failure. At the conclusion of the procedure,
contact Galil Medical Customer Service.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Reason for Appearance:
Fan X was identied as not operating.
Solutions:
After completing the procedure, contact
Galil Medical Customer Service.