EasyManua.ls Logo

BTG ICEfx - Page 89

BTG ICEfx
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ICEfx™ Cryoablation System
8-17
Message Reason for Appearance / Solutions
40-05 Used Needle
This needle has been previously used
Replace with a new needle
Reason for Appearance:
The software prevents needle reuse and
identied a needle previously used. The
channel will be disabled until a new needle
is connected.
Solutions:
1. Disconnect and segregate the needle.
2. Replace with a new needle.
40-06 Corrupt Memory
Needle Memory chip is corrupt on Channel X
Manually select the needle type
Reason for Appearance:
User connected a needle containing a bad
memory chip.
Solutions:
Manually select the needle type using the
drop-down menu.
40-07 Passive Thaw Required
The currently selected needle on channel [x] does
not have i-Thaw capability
Passive thaw is required on this channel
Reason for Appearance:
Combinations of needles that are
i-Thaw capable and that are not i-Thaw
capable were connected to the needle
connection panel.
OR
Passive thaw is required.
Solutions:
Wait for ice to passively thaw.
40-08 i-Thaw Error
Channel [Xa or Xb] i-Thaw needle is defective for
electrical thawing
Use passive thaw
Reason for Appearance:
An i-Thaw needle was identied as not
i-Thaw operational.
Solutions:
Use passive thawing.

Table of Contents