8-20 User Manual
Message Reason for Appearance / Solutions
60-09 Software Recovery
This will rollback the software on this system
Are you sure you want to do this?
YES NO
Reason for Appearance:
User pressed the Software Recovery
button then selected Rollback. Activation
will rollback the software to the previous
software version.
Solutions:
Conrm software rollback. Reference
Section 8.1 for information on software
recovery.
60-10 Software Recovery
Are you sure you want to restore the default
settings for all congurations?
YES NO
Reason for Appearance:
User pressed the Software Recovery
button then selected Load. Activation will
restore the system default settings for all
congurations.
Solutions:
Conrm restoration of default settings.
60-11 Incompatible Hardware
The hardware is not compatible with the current
software
Try downloading a software update from the Galil
Medical servers or from a Galil Medical USB ash
drive
Reason for Appearance:
When the system performed self-checks
at start up, an incompatibility between the
hardware and software was detected.
Solutions:
Download a software update from the
Galil Medical servers.
OR
Manually update the software using
the provided Galil Medical USB ash
drive.