EasyManua.ls Logo

BTG ICEfx - Page 92

BTG ICEfx
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-20 User Manual
Message Reason for Appearance / Solutions
60-09 Software Recovery
• This will rollback the software on this system
• Are you sure you want to do this?
YES NO
Reason for Appearance:
User pressed the Software Recovery
button then selected Rollback. Activation
will rollback the software to the previous
software version.
Solutions:
Conrm software rollback. Reference
Section 8.1 for information on software
recovery.
60-10 Software Recovery
• Are you sure you want to restore the default
settings for all congurations?
YES NO
Reason for Appearance:
User pressed the Software Recovery
button then selected Load. Activation will
restore the system default settings for all
congurations.
Solutions:
Conrm restoration of default settings.
60-11 Incompatible Hardware
• The hardware is not compatible with the current
software
• Try downloading a software update from the Galil
Medical servers or from a Galil Medical USB ash
drive
Reason for Appearance:
When the system performed self-checks
at start up, an incompatibility between the
hardware and software was detected.
Solutions:
• Download a software update from the
Galil Medical servers.
OR
• Manually update the software using
the provided Galil Medical USB ash
drive.

Table of Contents