SCPI Command
*STB?
It queries service request status register.
Query command format: *STB?
Return data: <NR1>
DISPlay Subsystem
DISPlay:STATe?
It queries the display location currently.
Parameter: None
Return data: LC Test / Contact Check / LC Null
DISPlay:LCTest
It sets function screen to LC Test mode.
Set command format: DISPlay:LCTest
Parameter: None
Return data: None
DISPlay:WVTest
It sets the function screen to WV Test mode.
Set command format: DISPlay:W VTest
Parameter: None
Return data: None
DISPlay:CONTact
It sets function screen to Contact Check.
Set command format: DISPlay:CONTact
Parameter: None
Return data: None
DISPlay:NULL
It sets function screen to LC Null.
Parameter: None
Return data: None
Set command format: DISPlay:NULL
ABORt Subsystem
ABORt
It aborts measurement procedure and enters DISCHARGE mode directly.
Set command format: ABORt
Parameter: None
Return data: None
TRIGger Subsystem
TRIGger:IMMediate
It follows the mode on the screen currently to start measurement procedure.
Set command format: TRIGger:IMMediate
Parameter: None