Battery Cell Insulation Tester 11210 User’s Manual
10-14
WVTest:MEASure:VMON?
It queries the 11210 output voltage at present.
Set command format: WVTest:MEASure:VMON?
Parameter: None
Return data: <NR2>, unit: V
CORRection Subsystem
CORRection:CONTact:STATe
It sets or queries Contact Check enabled state.
Set command format: CORRection:CONTact:STATe {<NR1>|ON|OFF}
Paramateer: <NR1> 0:OFF, 1:ON
Return data: <NR1> 0:OFF, 1:ON
CORRection:CONTact:NULL[:IMMediate]
It enables measurement Contact Check NULL.
Set command format: CORRection:CONTact:NULL
Parameter: None
Return data: <NR1>
CORRection:CONTact:DUT[:IMMediate]
It enables measurement Contact Check DUT.
Set command format: CORRection:CONTact:DUT
Parameter: None
Return data: <NR1>
CORRection:CONTact:VAL
It sets Contact Check VAL.
Set command format: CORRection:CONTact:VAL {<NR1>|MIN|MAX}
Parameter: <NR1>, None, MAX: 32768 acceptable, MIN:0
Return data: < NR1> no unit
CORRection:CONTact:DATA?
It queries Contact Check Null, DUT and calculated Contact Check judgment value (VAL).
Set command format: CORRection:CONTact:DATA?
Parameter: None
Return data: <NR1>, <NR1>, <NR1> NULL, DUT, VAL
CORRection:NULL:STATe
It sets or queries LC Null enabled state.
Set command format: CORRection:NULL:STATe {<NR1>|ON|OFF}
Parameter: <NR1> 0: OFF, 1: ON
Return data:<NR1> 0: OFF, 1: ON
CORRection:NULL
It conducts LC Null measurement.
Set command format: CORRection:NULL
Parameter: None
Return data: None
CORRection:NULL:DATA?
It queries measurement result of each LC Null range.
Set command format: CORRection:NULL:DATA?
Parameter: None