Battery Cell Insulation Tester 11210 User’s Manual
10-16
CALCulate:CONDition[:LCT]:UPPer:DATA
It sets or queries the upper limit of compare function.
Set command format: CALCulate:CONDition[:LCT]:UPPer:DATA {<NRf>|MIN|MAX}
Parameter: <NRF>, when IR mode the range is 1 ~ 1P(Ω) or
when LC mode the range is 1nA ~ 20mA
Return data: < NRF > The unit is in Ω when it is under IR mode and the unit is in mA
when it is under LC mode.
CALCulate:CONDition:PDV:ENABle
It sets or queries enabled compare PDV function.
Set command format: CALCulate:CONDition:PDV:ENABle {<NR1>|ON|OFF}
Parameter: <NR1> 0:OFF, 1:ON
Return data: <NR1> 0:OFF, 1:ON
CALCulate:CONDition:PDV:LEVEL:DATA
It sets or queries percentage threshold of PDV compare function.
Set command format: CALCulate:CONDition:PDV:LEVEL:DATA {<NR1>|MIN|MAX}
Parameter: <NR1>, Level, MAX:99 and MIN:1 are acceptable.
Return data: < NR1> Level
CALCulate:CONDition:PDI:ENABle
It sets or queries enabled compare PDI function.
Set command format: CALCulate:CONDition:PDI:ENABle {<NR1>|ON|OFF}
Parameter: <NR1> 0: OFF, 1: ON
Return data: <NR1> 0: OFF, 1: ON
CALCulate:CONDition:PDI:LEVEL:DATA
It sets or queries percentage threshold of PDI compare function.
Set command format: CALCulate:CONDition:PDI:LEVEL:DATA {<NR1>|MIN|MAX}
Parameter: <NR1>, Level, MAX:99 and MIN:1 are acceptable.
Return data: < NR1> Level
SYSTem Subsystem
SYSTem:DISCHarge:VOLTage
It sets or queries Discharge voltage threshold.
Set command format: SYSTem:DISCHarge:VOLTage {<NRf>|MIN|MAX}
Parameter: <NRF>, the range is from 0.5 to 10V, MAX: 10V and MIN: 0.5V are
acceptable.
Return data: < NRF > the unit is V.
SYSTem:ALARM
It sets or queries beeper alarm mode.
Set command format: SYSTem:ALARM <NR1>
Parameter: < NR1> 0 (PULSe) and 1 (CONTinuous) are acceptable.
Return data: 0 or 1
SYSTem:BEEPer:STATe
It sets or queries beeper volume.
Set command format:SYSTem:BEEPer:STATe {<NR1>|OFF|LOW|HIGH}
Parameter: < NR1>, 0 (OFF), 1 (LOW) and 2(HIGH) are acceptable.
Return data: < NR1>