SCPI Command
SYSTem:SOUND
It sets or queries measurement hint sound.
Set command format: SYSTem:SOUND {<NR1>}
Parameter: <NR1>, 0 and 1 are acceptable. 0 indicates to clear it to Low and 1
means to keep Hi-Z.
Return data: < NR1 >
SYSTem:CONTRast
It sets or queries LCD panel brightness.
Set command format: SYSTem:CONTRast <NR1|MIN|MAX >
Parameter: < NR1>, 1~15, MAX:15 and MIN:1 are acceptable.
Return data: < NR1>
SYSTem:LFREQuency
It sets or queries Line Frequency.
Set command format: SYSTem:LFREQuency {< NR1>|50|60}
Parameter: < NR1>, 0 (50) and 1 (60) (Hz) are acceptable.
Return data: < NR1>
SYSTem:HANDler
It sets or queries Handler setting parameter.
Set command format: SYSTem:HANDler {< NR1>|CLEAR|HOLD}
Parameter: < NR1>, 0 (CLEAR) and 1 (HOLD) are acceptable.
Return data: < NR1>
SYSTem:KLOCk
It sets or queries LCD key lock.
Set command format: SYSTem:KLOCk {<NR1>|OFF|ON}
Parameter: < NR1>, 0 (ON) and 1 (OFF) are acceptable.
Return data: < NR1> 0 or 1
SYSTem:DIGIts
It sets or queries displayed digit of measurement current (LC).
Set command format: SYSTem:DIGIts <NR1>
Parameter: < NR1>, 0 and 1 are acceptable. 0 indicates 3.5 digits and 1 means 4.5
digits.
Return data: < NR1> 0 or 1
SYSTem:RANGe:CHANGe:DELAY
It sets or queries range change delay time when auto ranging.
Set command format:
SYSTem:RANGe:CHANGe:DELAY {<NRf>|MIN|MAX}
Parameter: <NRF>, the range is from 0.005 to 9.999 s. MAX: 9.999s and MIN:
0.005s are acceptable.
Return data: < NRF >, the unit is S.
SYSTem:HANDler:CLEar:TIME
It sets or queries auto clear time of Handler state after measuring.
Set command format: SYSTem:HANDler:CLEar:TIME {<NRf>|MIN|MAX}
Parameter: <NRF>, the range is from 0.005 to 9.999 s, MAX: 9.999s and MIN:
0.005s are acceptable.
Return data: < NRF >, the unit is S.