Battery Cell Insulation Tester 11210 User’s Manual
5-8
(13) Alarm Mode:
The warning sound format needs to use with activated Beeper function and Sound
Mode setting while judging good and no good product. There are two modes of two
short sounds (Pulse) and one long sound (Level). The factory default value is Pulse.
Pulse: The warning sound sets to two short sounds once the tester judging good/no
good product.
Level: The warning sound sets to one long sound once the tester judging good/no
good product.
(14) Beeper
It sets the warning beep volume to OFF, LOW, and HIGH. The factory default setting
is LOW.
(15) Sound Mode:
The warning sound format contains two modes of PASS and FAIL when comparing
the measurement. The factory default setting is FAIL.
FAIL: It beeps when detected no good product during performing the measurement
of COMPARE.
PASS: It beeps when detected good product during performing the measurement of
COMPARE.
(16) LC Negative Current:
LC negative current display includes three modes of Linear, Zero and Abs. The
factory default is Abs.
Linear : It is the actual measurement current value with positive and negative
values for operation directly.
Zero : All negative current values are modified to 0 for operation.
Abs. : All negative current values take absolute value to modify to positive value
for operation.
(17) Hard Copy:
It saves PD Analyzer measured waveform or data with export format contains two
modes of PNG and CSV. The factory default is PNG. PD Analyzer function is
applied.
(18) Line Frequency:
It sets the line frequency to 50Hz or 60Hz for AC110V/220V power source. The
factory default is 50Hz.