EasyManua.ls Logo

Chroma 62000D Series - Page 144

Chroma 62000D Series
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programmable Bidirectional DC Power Supply 62000D Series
Operating and Programming Manual
5-18
SOURce:VOLTage:LIMIT:LOW <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:VOLT:LIMIT:HIGH 60.0 It sets the output voltage range
to 60V maximum.
SOUR:VOLT:LIMIT:LOW 20.0 It sets the output voltage range
to 20V minimum.
Query Syntax:
SOUR:VOLT:LIMIT:HIGH? [{MAX/MIN}]
SOUR:VOLT:LIMIT:LOW? [{MAX/MIN}]
Return Parameter:
<NRf+> [Unit=Volt]
Query Example:
SOUR:VOLT:LIMIT:HIGH? It returns the voltage high limit.
SOUR:VOLT:LIMIT:HIGH? MAX It returns the max. voltage can
be set.
Return Example:
8.000000e+01
3. SOURce:VOLTage:PROTect:HIGH
Description:
Sets the voltage range for over voltage protection.
Syntax:
SOURce:VOLTage:PROTect:HIGH <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:VOLT:PROT:HIGH 60.0 It sets the high limit to 60V for
voltage output protection.
Query Syntax:
SOUR:VOLT:PROT:HIGH? [{MAX/MIN}]
Return Parameter:
<NRf+> [Unit=Volt]
Query Example:
SOUR:VOLT:PROT:HIGH? It returns the high limit of
voltage protection.
SOUR:VOLT:PROT:HIGH? MAX It returns the max. over
voltage protection.
Return Example:
8.800000e+01
4. SOURce:VOLTage:SLEW
Description:
Sets the rising or falling slew rate (volt/ms) of output voltage.
Syntax:
SOURce:VOLTage:SLEW <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:VOLT:SLEW 0.01 It sets the output voltage slew rate
to 0.01 volt/ms.
SOUR:VOLT:SLEW 10 It sets the output voltage slew rate
to 10 volt/ms.
Query Syntax:
SOUR:VOLT:SLEW? [{MAX/MIN}]
Return Parameter:
<NRf+> [Unit=Volt/ms]
Query Example:
SOUR:VOLT:SLEW? It returns the voltage slew rate.
SOUR:VOLT:SLEW? MAX It return the max. voltage slew rate.
Return Example:
1.000000e+01
5. SOURce: CURRent
Description:
Sets the output current (ampere).
Syntax:
SOURce:CURRent <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR 1 It sets the output current to 1 amps.
SOUR:CURR 60.00 It sets the output current to 60.00 amps.
Query Syntax:
SOUR:CURR? [{MAX/MIN}]
Return Parameter:
<NRf+> [Unit=Amp]
Query Example:
SOUR:CURR? It returns the current.
SOUR:CURR? MAX It returns the max. current can be set.
Return Example:
1.000000e+00

Table of Contents