OperationModel 2500 143
© 2015-2017 Distek, Inc. All Rights Reserved
The dissolution bath will automatically turn off the heat when calibration is
being performed.
If the TempChek is not connected to the instrument, a popup warning message
will appear informing the user that "TempChek is not connected". Refer to TCS
Calibration Using Reference.
3 The
TCS Ca lib ra tion
screen is displayed as shown.
TCS Calibration
4 Depending on the unit and type of calibration, follow the procedure for either calibration
using a TempChek or Reference on the next chapters.