TroubleshootingModel 2500 190
© 2015-2017 Distek, Inc. All Rights Reserved
7.7 Temperature Calibration
Using TempChek
This section gives an overview of possible issues when calibrating the shaft sensors using a Distek
TempChek on the Model 2500 Dissolution System. To ensure consistent performance, follow the
recommended possible solutions given for a specific problem.
Problem
Possible Solution
N o displa y from
Temp C hek
N o response from
Temp C hek
TempChek
Check the battery (9 volt) of the TempChek. Replace if necessary.
RS-232
Check the serial cable integrity. Replace if necessary.
Single-Board
Computer
Check the internal serial cable integrity and connection. Check
the orientation of the internal serial cable going to the single
board computer (SBC).
Faulty single board computer (SBC). Replace if necessary.