EasyManua.ls Logo

Distek 2500 - Page 151

Distek 2500
206 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OperationModel 2500 151
© 2015-2017 Distek, Inc. All Rights Reserved
4 Once the temperature has equilibrated, using the number pad, enter in the reference
temperature into the
Ref. Tem p. °C
parameter
then quickly select
Ca libra te
to start vessel
1 calibration as shown.
Select Calibrate to calibrate vessel 1
5 Observe that the
V essel Tem p °C
closely matches (±0.05°C) the reference current
temperature reading and that the
Lo-C a l
parameter is updated. Repeat steps 4 and 5 if it
does not meet the criteria.
6 Once satisfied with the result, continue the calibration by selecting vessel position 2 and
repeat steps 3 to 5.
7 Repeat the above steps for the remaining vessel positions.
8 Once the Lo-Cal is complete, exit the shaft calibration screen by selecting
Done
.
9 Set the bath temperature to 41°C and wait for the vessel temperature to equilibrate.
10 Repeat steps 1 to 5 to start the
Hi-Ca l
calibration.
11 Observe that the
V essel Tem p °C
closely matches (±0.05°C) the reference current
temperature reading and that the
Hi-Ca l
parameter is updated. Repeat steps 4 to 11 if it
does not meet the criteria.
12 Repeat the above steps for the remaining vessel positions.
13 Once the Hi-Cal is complete, exit the shaft calibration screen by selecting
Done
.

Table of Contents

Other manuals for Distek 2500

Related product manuals