If vibration pads are used underneath the circulator, the pads should not be more than
1/2 inch (1.27cm) thick, otherwise the pump will not self prime.
6 Properly secure the tubing using the hose clamps and a pair of pliers to tighten the clamps.
The acrylic bath is susceptible to cracking/crazing when cleaned with chemicals
such as alcohol (isopropyl), acetones and ammonia based cleaners and therefore
should mainly be cleaned with mild soap and water.
Make sure that the ventilation holes of the thermocirculator are not
obstructed. A clearance of 3 inches is advisable.
Clean the inside of the bath for any packaging material that might lodge inside
and damage the pump during normal operation.
The pump is self priming so there is no need to prime the TCS.
Do not let the pump run dry.