OperationModel 2500 99
© 2015-2017 Distek, Inc. All Rights Reserved
5 Another status window prompts the operator to perform the following:
Raise the drive head
Attach the baskets (with dosage)
Lower the drive head and click
Sta rt
to start timer test.
Verify, raise drive and introduce dosage
If using the countdown timer in the dialog window, lower the drive
head when the timer reaches between 3 to 5 seconds. This will
allow the stirring to start automatically.
6 Testing starts. Elapsed Time and Sampling (All
)
windows are updated.
Test started
7 Once the test reaches 20 seconds before the first sampling interval, the sampling time will
be displayed in seconds, prompting the operator to get ready to collect the samples
manually. Collect all samples at zero seconds.
This step will repeat if the method is programmed to have multiple sampling intervals.