Reprocessing
106 Instructions for use Carina SW 3.2n
Dismantling
Before dismantling
1 Switch off the device and any additional devices
and disconnect their power plugs.
2 Drain the water traps and breathing hoses.
3 Drain the water container of the breathing gas
humidifier.
Removing breathing hoses
1 Remove the expiratory valve or leakage valve
as required. Remove the pilot line of the expira-
tory valve.
2 Remove the breathing hoses. Drain conden-
sate from the breathing hoses.
3 If fitted: Pull the water trap from the breathing
hoses.
4 Pull the water trap container from the water
trap. Drain the water trap.
5 Pull the bacterial filter from the inspiratory port.
Dispose of disposable products and reprocess
reusable breathing hoses in accordance with
the corresponding instructions for use.
Dismantling accessories
Dismantle and reprocess the breathing gas hu-
midifier and the medication nebulizer in accord-
ance with the corresponding instructions for
use.
Dismantle and dispose of the bacterial filter and
HME in accordance with the corresponding in-
structions for use.