Reprocessing
108 Instructions for use Carina SW 3.2n
Classifications for reprocessing
Classification of medical devices
Medical devices and their components are classi-
fied according to the type of application and the risk
resulting from it.
Classification of the device-specific components
Observe the instructions for use for the compo-
nents.
Non-critical
– Carina basic device
– Trolley with accessories
–O
2 supply hose
– Hinged arm
– Bed and wall mountings
Semi-critical B
– Reusable water trap
Classification Definition
Non-critical Components that merely come into contact with intact skin.
Semi-critical Components that carry breathing gas or come into contact with mucous mem-
branes or pathologically altered skin.
Semi-critical A No special requirements regarding reprocessing
Semi-critical B Increased requirements regarding reprocessing
Critical Components that penetrate skin or mucous membranes, or come into contact
with blood