Technical data
136 Instructions for use Carina SW 3.2n
Alarm delays
Alarms are not displayed immediately when the
alarm limit is exceeded.
Alarm Delay time
Airway pressure high !!! If the alarm limit is exceeded in 2 consecutive breathing phases
Airway pressure low !!! >Tdiscon + 5 s
Apnea !!! After Tapn has elapsed, if a time has been set for Tapn
Apnea Ventilation ! After Tapn has elapsed, if a time has been set for Tapn
Breath. circuit incorrect !!! >37 s
Continuous Paw high !!! >15 s
Continuous Paw low !!! >30 s (at least 3 breaths)
HPO supply insufficient !!! >30 s
Large leakage ! If the permissible patient leakage is exceeded from the 11th breathing
phase
LPO pressure high !! If the alarm limit is exceeded in 2 consecutive breathing phases
MV low !!! 60 s at start of ventilation and after reset of the alarm
Airway pressure low !!!
RR high !!! 60 s