Technical data
124 Instructions for use Carina SW 3.2n
Tidal volume with apnea ventilation VTapn
Range 100 to 2000 mL, BTPS
Resolution 10 mL
Volume Guarantee VG
Range On, Off
Inspiratory O
2 concentration (in HPO mode) FiO2
Range 21 to 100 %
Resolution 1 %
AutoAdapt Aut.Adapt
Range On, Off
Positive end-expiratory pressure with AutoAdapt PEEPadp
Range 3 to 20 mbar (or hPa or cm H
2O)
Resolution 1 mbar (or hPa or cm H
2O)
Pressure support with AutoAdapt
Δ
PSadpt
Range 0 to 37 mbar (or hPa or cm H2O)
Resolution 1 mbar (or hPa or cm H
2O)
Time in AutoAdapt Tadpt
Range 5 to 120 min
Resolution 1 min
Performance characteristics
Control principle Time-controlled or flow-controlled and pressure-
controlled or volume-controlled
Supply system for patient flow Turbine
Maximum continuous inspiratory flow 180 L/min, BTPS (of which max. 120 L/min with
100 % O2)
Device resistance during device failure
The resistance is specified for the breathing circuit,
i.e., between the emergency breathing valve (un-
derpressure), the expiratory valve, and the patient
connection. The specified resistances take the
breathing gas humidifier in account where neces-
sary but not additional accessories such as bacte-
rial filters or HMEs. See the applicable instructions
for use for these components for the resistance.
Setting values (cont.)